RPTLoad version 2 PC Software is the replacement for RPTLoad version 1 (old DHI freeware) and is used to write or read Fluke “RPT” sensor characterization information stored within a product or sensor using the RS232 port.
It works with legacy (old) products and newer models like PM600 and PM630 measure modules. The old version of RPTLoad does not work with PM Modules.
Download the below zip file and read the file named "RptLoad2 for Auth Svc Providers 20200413.docx"
Note that there are multiple versions of coefficient files. There always have a .V01 extension and always start with the serial number of the transducer. For example SN104342.V01
Newer coefficient files can end with .V01, _P.v01, _P!.v01,_A.v01, or _A!.vo1
The below matrix shows which file to use with different models of equipment.
To choose the file to upload to the device, leave the "Serial Number to load: " entry box blank. If you type a serial number in this entry box, then RPTLoad will search a particular folder for the .V01 file (and may not find it).
Choose the menu path: Setup > Upload from File
Then navigate to the location with the .V01 file that you want to upload to the device.
Wait for software to load the file. The “Header Information” and “Coefficient Information” should have values show up.
When then file is fully loaded a message at the bottom will appear that says “Coefficients written to RPT successfully”.