Tech Notes
--------------------------------------------------------------------------------------------------------------------------------
Title: 5606 probe accuracies at -197°C, -38°C, and 100°C
Created: 8 Apr 2013
Last Revised:
--------------------------------------------------------------------------------------------------------------------------------
Applies To: 5606
Problem Description: How can I find or determine the accuracies on the 5606?
Resolution/Work Around: The accuracies can be determined by the following:
| 
 
  | 
 -197°C  | 
 -38°C  | 
 100°C  | 
| 
 drift  | 
 0.0057  | 
 0.0252  | 
 0.0417  | 
| 
 cal uncertainty  | 
 0.025  | 
 0.025  | 
 0.045  | 
| 
 Sum (RSS)  | 
 0.000657  | 
 0.00126  | 
 0.003764  | 
| 
 Accuracy °C=  | 
 0.026  | 
 0.035  | 
 0.06  | 
		