Tech Notes
--------------------------------------------------------------------------------------------------------------------------------
Title: Factory PID settings for my Metrology Well
Created: 9 Jan 2013
Last Revised:
--------------------------------------------------------------------------------------------------------------------------------
Applies To: Various Metrology Wells
Problem Description: Unit will not maintain or reach temperature. PID is wrong.
Resolution/Work Around: Set the PID back to factory default.
| 
 4180  | 
|
| 
 Temp PBand  | 
 6.0  | 
| 
 Temp Int  | 
 30.0  | 
| 
 Temp Der  | 
 3.0  | 
| 
 Emissivity  | 
 0.95  | 
| 
 4181  | 
|
| 
 Temp PBand  | 
 10.0  | 
| 
 Temp Int  | 
 30.0  | 
| 
 Temp Der  | 
 5.0  | 
| 
 Emissivity  | 
 0.95  | 
| 
 9142  | 
|
| 
 Temp PBand  | 
 2.0  | 
| 
 Temp Int  | 
 30.0  | 
| 
 Temp Der  | 
 1.0  | 
| 
 9143  | 
|
| 
 Temp PBand  | 
 18.0  | 
| 
 Temp Int  | 
 90.0  | 
| 
 Temp Der  | 
 30.0  | 
| 
 9144  | 
|
| 
 Temp PBand  | 
 18.0  | 
| 
 Temp Int  | 
 90.0  | 
| 
 Temp Der  | 
 30.0  | 
| 
 9170  | 
|
| 
 Temp PBand  | 
 1.5  | 
| 
 Temp Int  | 
 20.0  | 
| 
 Temp Der  | 
 1.5  | 
| 
 Grad PBand  | 
 5.0  | 
| 
 Grad Int  | 
 40.0  | 
| 
 9171  | 
|
| 
 Temp PBand  | 
 3.0  | 
| 
 Temp Int  | 
 40.0  | 
| 
 Temp Der  | 
 1.5  | 
| 
 Grad PBand  | 
 10.0  | 
| 
 Grad Int  | 
 80.0  | 
| 
 9172  | 
|
| 
 Temp PBand  | 
 12.0  | 
| 
 Temp Int  | 
 90.0  | 
| 
 Temp Der  | 
 10.0  | 
| 
 Grad PBand  | 
 6.0  | 
| 
 Grad Int  | 
 50.0  | 
| 
 Grad Ratio  | 
 1.4  | 
| 
 9173  | 
|
| 
 Temp PBand  | 
 6.0  | 
| 
 Temp Int  | 
 100.0  | 
| 
 Temp Der  | 
 10.0  | 
| 
 Grad PBand  | 
 10.0  | 
| 
 Grad Int  | 
 150.0  | 
| 
 Grad Ratio  | 
 1.3  | 
| 
 9190A  | 
|
| 
 Temp PBand  | 
 10.0  | 
| 
 Temp Int  | 
 100.0  | 
| 
 Temp Der  | 
 8.0  | 
| 
 Grad PBand  | 
 20.0  | 
| 
 Grad Int  | 
 150.0  | 
		